Laparoscopic instrument tracking systems are a key element in image-guided interventions, which requires high accuracy to be used in a real surgical scenario. In addition, these systems are a suitable option for objective assessment of laparoscopic technical skills based on instrument motion analysis. This study presents a new approach that improves the accuracy of a previously presented system, which applies an optical pose tracking system to laparoscopic practice. A design enhancement of the artificial markers placed on the laparoscopic instrument as well as an improvement of the calibration process are presented as a means to achieve more accurate results. A technical evaluation has been performed in order to compare the accuracy between the previous design and the new approach. Results show a remarkable improvement in the fluctuation error throughout the measurement platform. Moreover, the accumulated distance error and the inclination error have been improved. The tilt range covered by the system is the same for both approaches, from 90° to 7.5°. The relative position error is better for the new approach mainly at close distances to the camera system.
机译:腹腔镜仪器跟踪系统是图像引导干预中的关键元素,需要在真正的手术场景中使用高精度。此外,这些系统是基于仪器运动分析的腹腔镜技术技能的客观评估的合适选择。本研究提高了一种提高了先前呈现的系统的准确性的新方法,该系统将光学姿势跟踪系统应用于腹腔镜实践。呈现在腹腔镜仪器上的人工标记的设计增强以及校准过程的改进作为实现更准确的结果的方法。已经进行了技术评估,以比较先前设计与新方法之间的准确性。结果在整个测量平台上显示了波动误差的显着改进。此外,已经提高了累积距离误差和倾斜误差。两个方法覆盖的倾斜范围对于两种方法相同,从90°到7.5°。相对位置误差主要用于新方法,主要是靠近相机系统的距离。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号