Comparison of Polar V800 with Healink R211B for the Purpose of Analyzing Heart Rate Variability

机译:HEALINK R211b对HEALINK R211B的偏光V800的比较,以分析心率变异性

获取原文

摘要

Heart rate variability (HRV) is generally accepted as the best method to evaluate the function of the human autonomic nervous system.This paper compares the differences between the Polar V800 and the Healink R211B in the application of HRV analysis.Twenty-nine healthy college students participated in the study.The data were recorded simultaneously by both devices from all the subjects in a resting state.After being processed,the RR interval (RRI) data for 26 of the cases from the two devices were imported into the Kubios HRV software and were approximately consistent in regards to the HRV parameters.There was no significant difference (r > 0.99,p > 0.05) between the two devices except for the al parameter (p = 0.021).For the other 3 cases,the Polar V800 RRI could not be corrected due to partially missing data.Therefore,the HRV analysis and comparison were not conducted.The results indicate that the Polar V800 and the Healink R211B combined with Kubios HRV software can be used for HRV analysis in the time domain,frequency domain and for partial nonlinear parameter analysis,except in cases where data were missed by the Polar V800.However,it should be noted that approximately 10% of the samples from the Polar V800 have more RRI values missing so the entire sample data could not be used for HRV analysis.
机译:心率变异性(HRV)通常被认为是评估人类自主神经系统功能的最佳方法。本文比较了极地V800与Healink R211B之间的差异在HRV分析中的应用中.Twenty-九个健康大学生参加了该研究。数据通过来自休息状态的所有拍摄对象的两个设备同时记录。正在处理的后,从这两个设备的26个案例的RR间隔(RRI)数据被导入Kubios HRV软件和关于HRV参数大致一致。除了AL参数之外,两种器件之间没有显着差异(R> 0.99,p> 0.05)(P = 0.021)。对于其他3例,极性V800 RRI可以由于部分缺失的数据不被纠正。因此,没有进行HRV分析和比较。结果表明,Polar V800和Headink R211b与Kubios HRV软件相结合可用于HRV A在时域,频域和部分非线性参数分析中的缺乏分析,除了由极性V800错过数据的情况。然而,应该注意的是,来自极性V800的大约10%的样本具有更多的RRI值缺失整个样本数据不能用于HRV分析。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号