首页> 外文会议>National Conference on Biomedical Engineering >Correcting Electromagnetic Tracker's Position Error Using Orientation Information Based on Hardy's Multi-Quadric Method
【24h】

Correcting Electromagnetic Tracker's Position Error Using Orientation Information Based on Hardy's Multi-Quadric Method

机译:基于Hardy的多反流方法校正电磁跟踪器的位置错误

获取原文

摘要

Electromagnetic tracking (EMT) systems are often used to track the position and orientation of a navigation tool during navigated surgery operations. But, their sensitivity to field distortions originating from electromagnetic emitting devices or metallic objects present in the environment reduces measurement accuracy. The purpose of this study is to incorporate orientation information in correcting position data provided by EMT based on Hardy's Multi-Quadric method (HMQ). We implemented the HMQ algorithm in two ways: the basic form in which location data is used only and the modified form in which both the orientation and location data are incorporated into a normalized vector in the relevant formula. Comparing the corrected position data with their corresponding reference values, which are provided by the optical tracking system, it is observed that the mean global error is reduced approximately by half and the variance is decreased nearly to its one tenth.
机译:电磁跟踪(EMT)系统通常用于跟踪导航手术操作期间导航工具的位置和取向。但是,它们对源自环境中存在的电磁发射器件或金属物体的场畸变的敏感性降低了测量精度。本研究的目的是基于Hardy的多反流方法(HMQ)来合并EMT提供的校正位置数据的方向信息。我们以两种方式实现了HMQ算法:仅使用位置数据的基本形式以及所修改的形式,其中取向和位置数据都结合到相关公式中的归一化向量中。将校正位置数据与光学跟踪系统提供的相应参考值进行比较,观察到平均全局误差大约减少一半,并且方差几乎减小到其十分之一。