To investigate the hemodynamic performance of overlapping bare-metal stents treatment of thoracic aortic aneurysms (TAA), three simplified geometric models, named, no stent, with one single stent and two overlapped stents were studied and compared in terms of velocity profiles and wall shear stress (WSS) distributions by means of computational fluid dynamics (CFD). The results showed that overlapping bare-metal stents intervention significantly decreased the WSS and the blood flow velocity in the aneurysm sac, meanwhile the vortex in the sac became weaker and weaker. The results indicated that overlapping stents intervention may effectively isolate the thoracic aortic aneurysm, protecting it from rupture. In conclusion, overlapping bare-metal stents may serve a similar purpose to that of the multilayer flow modulator (MFM) manufactured by Cardiatis SA (Isnes, Belgium).
机译:为了研究重叠裸金属支架的血流动力学性能治疗胸主动脉瘤(TAA),研究了三种简化的几何模型,并在速度剖面和墙面剪切方面进行了比较,并比较了一个单支架和两个重叠支架。通过计算流体动力学(CFD)的应力(WSS)分布。结果表明,重叠的裸金属支架干预显着降低了动脉瘤SAC中的WSS和血流速度,同时囊中的涡流变弱,较弱。结果表明,重叠的支架干预可以有效地分离胸主动脉瘤,保护其免受破裂。总之,重叠的裸金属支架可以为Cardiatis SA(isnes,比利时)制造的多层流动调制器(MFM)的类似目的。