Establishing tumour tracking accuracy in free-breathing respiratory gated SBRT of lung cancer

机译:在肺癌的自由呼吸呼吸道门静脉中建立肿瘤跟踪准确性

获取原文

摘要

Free-breathing respiratory gated SBRT of surgically inoperable lung cancer has been clinically commissioned. This study was to establish the tumour tracking accuracy under clinical conditions based on an implanted fiducial marker. A Visicoil? marker embedded in tissue-equivalent material mounted in a phantom (ET Gating Phantom? Brainlab) driven by a patient's breathing data was treated with the ExacTrac? system. This one-dimensional moving marker represented a tumour motion in superior-inferior (S-I) direction measured through 4DCT study of the same patient. Both Gafchromic? films and the stereoscopic kV images were used for tracking the position of the marker. For tumour motion at magnitudes of 10, 20 and 29 mm and treated with corresponding gate widths of 50%, 33% and 20% of free breathing amplitude. the implanted marker was able to be tracked with a deviation <1.53 mm to its planned position.
机译:自由呼吸呼吸门控SBRT的手术不可操作的肺癌已在临床上委托。本研究是基于植入基准标记的临床条件下建立肿瘤跟踪精度。粘纸?用患者呼吸数据驱动的患有患者(ET浇口幽灵般的组织 - 等效材料中的标记用患者的呼吸数据驱动,用Exactrac治疗?系统。该一维移动标记物代表了通过同一患者的4DCT研究测量的优异(S-I)方向上的肿瘤运动。 Gafchromic都是什么?薄膜和立体kV图像用于跟踪标记的位置。对于10,20和29mm的幅度下的肿瘤运动,并用相应的栅极宽度为50%,33%和20%的自由呼吸振幅处理。能够以偏差<1.53mm达到其计划位置跟踪植入标记。