首页> 外文会议>IEEE Nuclear Science Symposium >The New Read-Out Electronics for the BaF{sub}2-Calorimeter TAPS
【24h】

The New Read-Out Electronics for the BaF{sub}2-Calorimeter TAPS

机译:BAF {Sub} 2 - 热计点水龙头的新读出电子产品

获取原文

摘要

A highly compact and fast VME based read-out board for BaF{sub}2 scintillation detectors has been designed, developed and finally tested in an in-beam experiment. Adapted to the excellent properties of BaF2, the unit allows to digitize time, energy and pulse-shape information of 4 detector channels in parallel. The board is piggy-back plugged onto a motherboard containing a high-speed 12-bit ADC and the VME Interface, commercially available in the customized version CAEN V874A. Both together combine to one single VME slot. A first measurement of the photon response of a TAPS sub-array with energy marked photons up to 2.6 GeV has documented the full functionality and excellent performance.
机译:为BAF {Sub} 2闪烁探测器的高度紧凑且快速的VME读出板已经设计,开发,并最终在梁梁实验中进行了测试。适用于BAF2的优异性能,该装置允许并行地将4个检测通道的时间,能量和脉冲形状信息进行数字化,能量和脉冲形状信息。板载捎带插入包含高速12位ADC和VME接口的主板,可在“定制版本CAEN V874A中”。两者都组合在一个单一的VME插槽中。第一次测量点击子阵列的光子响应具有高达2.6 GEV的能量标记的光子,记录了完整的功能和优异的性能。