Calibration method for creating an inter-instrumental and inter-laboratory spectral databank

机译:用于创建仪器间和实验室间光谱数据库的校准方法

获取原文

摘要

Different trials in different instruments and laboratories were performed with and without tune compound-based procedures. From the experimental results, it was shown that the tune method seems to be a requirement for inter-laboratory exchange. In the case of comparison of CID spectra from tandem in time, it appeared that the tune method was not crucial. However, the inter-instrumental comparison without application of a similar tune method demonstrates the requirement of rigorous experimental condition to obtain the same spectral fingerprint between tandem in time and tandem in space. Some ions may also be favored because of the internal energy deposited and the type of excitation used (resonant or non-resonant). As expected, the consecutive dissociations were favored in tandem with "axis collisions" and these consecutive fragments cannot be seen or they are appeared in weak abundance for tandem with temporal sequences. Instrumental characteristics indicate the limits of the tune procedure to improve the comparison of CID spectra recorded with tandem with "axis collisions" and tandem with temporal sequences.
机译:不同仪器和实验室的不同试验在没有基于曲调的基于化合物的程序进行。从实验结果来看,表明曲调方法似乎是实验室交流的要求。在随着时间的推移与CID光谱比较的情况下,似乎曲调方法并不至关重要。然而,不应用类似曲调方法的局部比较表明了严格的实验条件的要求,以获得串联之间的相同光谱指纹和空间串联。由于沉积的内部能量和使用的激发类型(共振或非谐振),一些离子也可能受到青睐。正如预期的那样,连续的解离在串联中有利于“轴碰撞”,并且不能看到这些连续的片段,或者它们出现在伴时间序列的扭曲弱丰富。仪器特性表示调谐程序的限制,以改善记录的CID光谱与“轴碰撞”和串联串联记录的CID光谱比较。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号