首页> 外文会议>American Society for Mass Spectrometry Conference on Mass Spectrometry and Allied Topics >Optimization of Negative Ion Chemical Ionization Mass Spectrometry for Explosives for Two GC/MS Instruments from Different Manufacturers
【24h】

Optimization of Negative Ion Chemical Ionization Mass Spectrometry for Explosives for Two GC/MS Instruments from Different Manufacturers

机译:不同厂商两种GC / MS仪器爆炸物的负离子化学电离质谱优化

获取原文

摘要

Varying reagent/collision gas flow rates (source pressures) yielded similar effects on explosives sensitivities for the two instruments. Source temperatures affected both sensitivities of explosives as well as their mass spectra for both instruments. Optimal sensitivities ocurred at different source temperatures for the two instruments. Higher source temperatures consistently shifted relative abundances to lower m/z fragment ions. The optimal injector temperature for instrument 2 was determined to be 250 deg C. Injector temperatures above this optimum resulted in complete degradation of some of the energetic compounds, while lower temperatures yielded insufficient volatilization of the explosives and related compounds.
机译:不同的试剂/碰撞气体流量(源压力)对两种仪器的爆炸物敏感性产生了类似的效果。源极温度影响爆炸物的敏感性以及两种仪器的质谱。两种仪器不同源温度造成的最佳敏感性。较高的源极温度始终如连地移动到下部M / Z片段离子的相对丰度。仪器2的最佳喷射器温度被确定为250℃的温度高于该最佳温度,导致一些能量化合物的完全降解,而较低的温度会产生爆炸物和相关化合物的不充分挥发。