Methods in vogue for the measurement of venous refilling time (VRT), based on the photo-plethysmo-graphic (PPG) principle, employ a feedback compensation technique to remove the influence of the colour of a patient's skin. Hence these methods require a "standardization phase" for every measurement. Based on an appropriate model for the propagation of light through skin and tissue, a novel signal processing method is proposed for the measurement of VRT. The proposed technique renders the measurement of VRT independent of not only the colour of the skin of a patient but also of source intensity and detector sensitivity. Test results obtained from a prototype unit, built under the LabVIEW virtual environment, establish the validity of the model and the proposed signal processing technique.
机译:用于测量静脉再填充时间(VRT)的时尚的方法,基于光解倍差 - 图形(PPG)原理,采用反馈补偿技术来消除患者皮肤颜色的影响。因此,这些方法需要每次测量的“标准化阶段”。基于用于通过皮肤和组织的光传播的适当模型,提出了一种用于测量VRT的新型信号处理方法。所提出的技术使VRT的测量变得独立于患者皮肤的颜色而且源强度和检测器灵敏度的颜色无关。从LabVIEW虚拟环境下构建的原型单元获得的测试结果,建立了模型的有效性和所提出的信号处理技术。