A novel signal processing method for the measurement of venous refilling time

机译:一种测量静脉补充时间的新型信号处理方法

获取原文

摘要

Methods in vogue for the measurement of venous refilling time (VRT), based on the photo-plethysmographic (PPG) principle, employ a feedback compensation technique to remove the influence of the colour of a patient's skin. Hence these methods require a ldquostandardization phaserdquo for every measurement. Based on an appropriate model for the propagation of light through skin and tissue, a novel signal processing method is proposed for the measurement of VRT. The proposed technique renders the measurement of VRT independent of not only the colour of the skin of a patient but also of source intensity and detector sensitivity. Test results obtained from a prototype unit, built under the LabVIEW virtual environment, establish the validity of the model and the proposed signal processing technique.
机译:基于光体积描记法(PPG)原理,用于测量静脉补充时间(VRT)的流行方法采用反馈补偿技术来消除患者皮肤颜色的影响。因此,这些方法每次测量都需要一个“标准化阶段”。基于适当的光通过皮肤和组织传播的模型,提出了一种用于测量VRT的新型信号处理方法。所提出的技术使VRT的测量不仅与患者皮肤的颜色无关,而且与光源强度和检测器灵敏度无关。从在LabVIEW虚拟环境下构建的原型单元获得的测试结果确定了模型的有效性以及所提出的信号处理技术。