Development of Evaluation Test Method for the Possibility of Central Venous Catheter Perforation Caused by the Insertion Angle of a Guidewire and a Dilator

机译:由导丝插入角度和扩张器的插入角度引起的中央静脉导管穿孔的可能性的评价试验方法的研制

获取原文

摘要

The Seldinger technique is a well-established medical procedure to insert an indwelling device into blood vessel. Although the method is widely performed, various adverse effects due to the procedure have been reported. In order to improve safety, some hospitals originally compiled important reminders based on their experiences in an operation manual. The aim of our research is to provide evidences supporting failure behaviors with a mock system to promote awareness in clinical practice. This paper presents results of a newly developed test method to evaluate the possibility of perforation while inserting the dilator to the cervical vein after the guidewire placement. A tensile tester (AGI-250kN, Shimadzu Corp.) equipped with a load cell (SLBL-50N, Shimadzu Corp.) was utilized to measure the insertion force of a dilator to vein. The porcine vein was prepared to be aligned with 15mm and to be cut opened to expose the medial wall, and it was mounted on a custom-made device. A guidewire and a dilator were vertically suspended from the upper side of the tensile tester. During the dilator was approaching down to the vein through the guidewire at the constant velocity, the insertion force was measured and the tissue surface was observed simultaneously. The perforation force was examined in relation to the insertion angle from 15degrees to 60degrees. While veins were pressed only by a guidewire, perforation was not identified at all degrees. After inserting a dilator, veins were torn at 45degrees and at 60degrees more than ION. It was suggested that vessel wall could be thinner and be torn by multiple insertion. It is concluded that the usage situation of a guidewire and a dilator was assessed objectively and the angle-load relationship was quantified. The test would be helpful to propose safe approach for patients.
机译:Seldinger技术是一种良好的医疗程序,可将留置装置插入血管中。虽然该方法被广泛进行,但报道了由于该过程引起的各种不利影响。为了提高安全性,一些医院最初根据他们在操作手册中的经验编制了重要提醒。我们的研究目的是提供支持失败行为的证据,嘲笑系统促进临床实践的认识。本文提出了新开发的测试方法的结果,以评估穿孔的可能性,同时将扩张器插入导丝放置后将扩张器插入颈椎。配备有带有称重传感器(SLBL-50N,Shimadzu Corp.)的拉伸测试仪(AGI-250KN,Shimadzu Corp.)来测量扩张器的插入力静脉。制备猪静脉用15mm对准并被切割以暴露内侧壁,并且将其安装在定制的装置上。导丝和扩张器从拉伸测试仪的上侧垂直悬挂。在扩张器期间,在恒定速度下通过导丝向静脉接近静脉,测量插入力并且同时观察组织表面。检查穿孔力与从15度到60度的插入角度检查。虽然仅通过导丝被压制静脉,但在所有程度上都没有鉴定穿孔。在插入扩张器后,静脉撕裂在45℃下,超过离子60度。有人建议血管壁可能会稀释,并通过多重插入撕裂。得出结论,客观地评估导丝和扩张器的使用情况,并定量了角度载荷关系。该测试有助于提出患者的安全方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号