首页> 外文会议>International Mass Spectrometry Conference >Ultra-Sensitive Quantitation of Exenatide with Micro-Flow LC Trap-and-Elute and High Resolution and Triple Quadrupole Mass Spectrometry Workflow
【24h】

Ultra-Sensitive Quantitation of Exenatide with Micro-Flow LC Trap-and-Elute and High Resolution and Triple Quadrupole Mass Spectrometry Workflow

机译:用微流量LC陷阱和洗脱和高分辨率和三重四极杆质谱工作流程的超敏定量

获取原文

摘要

Comparing with the published best results where 100 pg/mL was determined as the LLOQ, 10 fold sensitivity improvement was observed on this? UHPLC-MSHRAM platform configured with nanoLC 425 and TripleTOF 5600. The LLOQ of this method was determined at 10 pg/ml with 50uL injection. Wide linear dynamic range was observed from 10 to 2,000 pg/mL with excellent precision (%CV < 15% throughout the linear range), and accuracy within 89.5% to 112%. The ultra-sensitive quantitation workflow does not suffer from throughput and the total run time for each sample is 10 minutes. Carryover was a known issue for exenatide quantitation. And in this study, trifluoroethanol was introduced to effectively remove residue exenatide from sampling path as well as from separation column to maintain low system background and to ensure low level carryover. Compared with regular wash, the wash mechanism utilized in this method could reduce carryover by 87.5% and the observed carryover was at 0.17%.
机译:与已发表的最佳结果相比,其中100pg / ml被确定为LLOQ,在此观察到10倍敏感性改善? UHPLC-MSHRAM平台配置有Nanolc 425和Tripletof 5600.该方法的LLOQ以50ul注射测定10pg / ml。宽线性动态范围从10至2,000pg / ml观察到具有优异的精度(在整个线性范围内的%CV <15%),并且精度在89.5%至112%以内。超敏感的定量工作流程不会遭受吞吐量,每个样本的总运行时间为10分钟。携带是杂交定量的已知问题。在这项研究中,引入三氟乙醇以有效除去从采样路径以及分离柱中除去残余物,以维持低系统背景并确保低水平的携带。与常规洗涤相比,该方法中使用的洗涤机制可降低87.5%,观察到的携带均为0.17%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号