首页> 外文会议>Conference on Computers in Cardiology >MRI in Patients with Implantable Devices: A Numerical Model for the Evaluation of Lead Heating
【24h】

MRI in Patients with Implantable Devices: A Numerical Model for the Evaluation of Lead Heating

机译:植入装置的MRI:铅加热评估的数值模型

获取原文

摘要

The radiofrequency (RF) field used in Magnetic Resonance Imaging (MRI) procedures may lead to significant increase in temperature and local specific absorption rate (SAR) for patients with implanted pacemaker (PM). In this paper two numerical models were developed to reproduce the heat generation process which occurs at the lead tip of an implanted PM when exposed to a RF field: the former (thermal model) shows how the position of temperature probes definitely affects the estimation of temperature rise and SAR value measured on the lead tip. The transversal contact between the active portion of the probe and the lead tip is the configuration associated with the highest estimation of temperature and SAR, whereas other configurations may lead to an underestimation close to 15% and 70% for temperature and SAR, respectively. The letter (RF model) predicted higher temperature increase and SAR value for right pectoral implants compared to left ones.
机译:磁共振成像(MRI)程序中使用的射频(RF)场可能导致植入起搏器(PM)患者的温度和局部特异性吸收率(SAR)的显着增加。在本文中,开发了两种数值模型来再现在暴露于RF场时植入PM的引线尖端发生的发热过程:前者(热模型)显示了温度探头的位置肯定会影响温度估计的情况在引线尖端测量的上升和SAR值。探针的有源部分与引线尖端之间的横向接触是与温度和SAR的最高估计相关的配置,而其他配置可以分别导致温度和SAR的低于15%和70%的低估。与左侧的字母(RF模型)预测右侧植入物的较高温度增加和SAR值。