首页> 外文会议>International Conference on Radionuclide Metrology and its Applications >Activity standardization of ~(3)H with the new TDCR system at PTB
【24h】

Activity standardization of ~(3)H with the new TDCR system at PTB

机译:PTB新型TDCR系统的活动标准化〜(3)小时

获取原文
获取外文期刊封面目录资料

摘要

A triple-to-double coincidence ratio (TDCR) liquid scintillation counting system (Pochwalski et al., 1988; Grau Malonda and Coursey, 1988) was set up at the Physikalisch-Technische Bundesanstalt (PTB) to be used as an additional technique for primary activity standardization. The optical chamber of the system was designed in our laboratory, expending great effort to achieve maximum light transmission to the photomultiplier tubes (PMT) and, consequently, a maximum counting efficiency and low uncertainties. A programme to compute the counting efficiencies and to analyse the measurement data was developed and tested. The software can also be used to allow for PMT asymmetries. The system and recently developed analysis procedures were used for the activity standardization of a tritiated water solution within the scope of an international comparison under the auspices of the Bureau International des Poids et Mesures (BIPM).
机译:在Physikalisch-Technische BundeSanstalt(PTB)中建立了三倍双倍的液体闪烁计数系统(Pochwalski等,1988; Grah Malonda和Carcory,1988),以用作额外的技术主要活动标准化。该系统的光学室在我们的实验室设计,满足了努力实现对光电倍增管(PMT)的最大光传输,从而最大限度地计数效率和低的不确定性。开发并测试了计算计数效率并分析测量数据的程序。该软件还可用于允许PMT不对称。该系统和最近开发的分析程序用于在国际局国际DES POIDS et Mesures(BIPM)的主持下的国际比较范围内的氚化水溶液的活动标准化。