首页> 外文会议>International Conference on Radionuclide Metrology and its Applications >Activity measurement of ~(85)Kr diluted by a large volume balloon technique
【24h】

Activity measurement of ~(85)Kr diluted by a large volume balloon technique

机译:大容量气球技术稀释的〜(85)KR的活性测量

获取原文

摘要

The National Metrology Institute of Japan (NMIJ) participated in the CCRI(II) Key-comparison of ~(85)Kr activity measurement. The absolute internal gas-counting method was used, with a set of proportional counters of different length. The original ~(85)Kr gas was diluted by a factor of 2 X 10~(5) in order to achieve the suitable activity concentration for the proportional counters. A new dilution method was introduced, based on a large-volume balloon and gravimetric determination of the dilution volume with small uncertainties.
机译:国家计量研究所(NMIJ)参加了CCRI(ii)键 - 比较〜(85)KR活性测量。使用绝对内部气体计数方法,其中一组比例计数器的不同长度。原始〜(85)KR气体稀释2×10〜(5)倍以实现比例计数器的合适活性浓度。基于大容量球囊和重量测定稀释体积,具有小的不确定性,引入了一种新的稀释方法。