Image Quality of a Photon-Counting Mammography System Compared to Digital Mammography Based on Amorphous Silicon with CsI-Scintillator

机译:基于CSI-Scintillator的非晶硅的数字乳房X线摄影系统的图像质量

获取原文

摘要

To compare image quality and dose of a photon-counting multi-slit scanner (PC) and a system based on amorphous silicon (aSi), images of the CDMAM 3.4 were taken in standard mode. For 3cm PMMA, the PC used 29kV/11.7mAs, aSi used 27kV/ 50mAs. For 5cm PMMA, PC used 35kV /14.8mAs, aSi used 31kV/ 50 mAs. Exposure was manually increased for PC and lowered for aSi-system. Average glandular dose and an image quality index (IQI) were calculated over the diameter ranges 0.06 - 2.0mm and 0.1-1. In standard mode with 3cm PMMA, IQI for PC was 35% lower than for aSi at 80% lower dose. Increased dose of PC resulted in 13% lower IQI at 57% lower dose. With 5cm PMMA IQI in standard mode was 18 % lower with PC at a 69% lower dose. Increasing the dose of PC resulted a 7% lower IQI at 54% lower dose. In conclusion the PC-system might reduce dose by up to 54% at equivalent image quality, although maximal quality of aSi could not be reached.
机译:为了比较光子计数的多缝扫描仪(PC)和基于非晶硅(ASI)的系统的图像质量和剂量,在标准模式下采用CDMAM 3.4的图像。对于3CM PMMA,PC使用29kV / 11.7mas,ASI使用27kV / 50mas。对于5CM PMMA,PC使用35kV /14.8mas,ASI使用31kv / 50 mas。 PC手动增加暴露并降低ASI系统。在0.06-2.0mm和0.1-1的直径范围内计算平均腺剂量和图像质量指数(IQI)。在具有3CM PMMA的标准模式下,PC的IQI低于80%低剂量的ASI的35%。增加的PC剂量导致13%的IQI降低剂量下降57%。使用5cm的PMMA IQI标准模式下降18%,PC下剂量为69%。增加PC剂量导致含量为7%的IQI,低剂量为54%。总之,PC系统在等效图像质量下可以将剂量减少到54%,尽管无法达到ASI的最大质量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号