首页> 外文会议>International Conference on Radiotherapy Gel Dosimetry >IMRT verification using a radiochromic/optical-CT dosimetry system
【24h】

IMRT verification using a radiochromic/optical-CT dosimetry system

机译:IMRT验证使用放射刻度/光学CT剂量测定系统

获取原文

摘要

The present work represents our first experiences relating to IMRT verification using a relatively new 3D dosimetry system consisting of a PRESAGE~(TM) dosimeter [3,4] (Heuris Inc, Pharma LLC) and an optical-CT scanning system (OCTOPUS~(TM) -^sMGS Inc). The work builds in a step-wise manner on prior work in our lab. The first step was a detailed investigating of the basic dosimetric properties of PRESAGE~(TM) on small volumes using an efficient in-house linear laser scanning system [4,5]. This was followed by application of a large volume PRESAGE dosimeter to verify a simple 3D dose-distribution [6], known to be well modelled by the planning system (ECLIPSE). Comparison of measured and planned distributions enabled evaluation of the PRESAGE~(TM)/OCTOPUS~(TM) dosimetry system. Confidence was further established by comparison with secondary independent measurement by EBT film. All these investigations confirmed the accurate performance of PRESAGE~(TM)/OCTOPUS~(TM) system, and provided a solid foundation to proceed to the present application: IMRT 3D dose verification.
机译:目前的作品代表了使用由概述〜(TM)剂量计[3,4](Heuris Inc,Pharma LLC)和光学-CT扫描系统(八达通_(八进)组成的相对较新的3D剂量系统与IMRT验证有关的第一经验。(章鱼〜( TM) - ^ SMGS INC)。工作在我们实验室的事先工作中以逐步的方式构建。第一步是使用高效内部线性激光扫描系统详细研究了小卷上的刻度〜(TM)的基本剂量性能[4,5]。然后施加大量刻度剂量计,验证简单的3D剂量分布[6],已知由规划系统(Eclipse)很好地建模。测量和计划分布的比较使得对刻录〜(TM)/八进孔〜(TM)剂量测定系统的评估。通过与EBT薄膜的二次独立测量相比,进一步建立了信心。所有这些调查确认了预设〜(TM)/八达通〜(TM)系统的准确性能,并为本申请提供了坚实的基础:IMRT 3D剂量验证。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号