首页> 外文会议>Annual Conference on Magnetism and Magnetic Materials >Erbium-doped gold sensor films for magnetic microcalorimeter x-ray detectors
【24h】

Erbium-doped gold sensor films for magnetic microcalorimeter x-ray detectors

机译:用于磁性微量电流仪X射线探测器的erbium掺杂的金传感器膜

获取原文

摘要

We briefly review the principles of a magnetic microcalorimeter x-ray detector and the requirements for fabricating an array of such sensors, which include a paramagnetic sensor fabricated with thin-film techniques. We discuss two methods for depositing an Er-doped Au sensor film and present magnetization measurements from 4 to 300 K in 5 T and near the detector's operating temperature and field. The properties of the deposited films that we have studied to date deviate from the expected behavior of bulk materials.
机译:我们简要介绍了磁微量仪X射线检测器的原理和制造这种传感器阵列的要求,包括用薄膜技术制造的顺磁传感器。我们讨论了两种用于将ER掺杂的AU传感器薄膜沉积的方法,并在5 T中将磁化测量值呈现为4至300 k,并在探测器的工作温度和场附近。我们已经研究过迄今为止的沉积薄膜的性质偏离了散装材料的预期行为。