Measurements of the residual stresses due to cement polymerization for cemented hip implants

机译:由于水泥聚合为粘合臀部植入物的残留应力的测量

获取原文
获取外文期刊封面目录资料

摘要

The initial fixation of the cemented hip prosthesis relies on the resistance of the interface between the metallic stem of the implant, the PMMA, and the adjacent bone. During the operation, the bone cement still in a liquid form is inserted between the implant and the bone. During polymerisation of the cement, residual stresses are generated in the bulk cement. An experiment has been devised to reproduce the in vivo behaviour of a cemented hip prosthesis, and to develop a technique to measure the residual stresses of the bone cement at the stem-cement interface. An idealized prosthesis (19-mm diameter) was placed inside a synthetic bone (outer diameter of 40 mm, inside diameter of 30 mm). The bone cement was poured between the stem and the bone. A sub-miniature load cell was inserted inside the idealized stem to measure directly the radial stresses generated by the cement curing on the hip stem. The tests are conducted at body temperature of 37°C to simulate the in-vivo conditions.
机译:粘合髋关节假体的初始固定依赖于植入物的金属茎与相邻骨之间的界面的电阻。在操作期间,仍然处于液体形式的骨水泥插入植入物和骨之间。在水泥的聚合过程中,在散装水泥中产生残余应力。已经设计了一种实验以再现巩固髋关节假体的体内行为,并开发一种测量茎水泥界面处的骨水泥残留应力的技术。将理想的假体(直径为19毫米)置于合成骨内(外径为40mm,内径为30mm)。骨水泥倒在茎和骨之间。将亚微型称重称为在理想化的杆内插入,以直接测量通过在髋部杆上的水泥固化产生的径向应力。测试在37℃的体温下进行,以模拟体内条件。