首页> 外文会议>International Symposium on Therapeutic Ultrasound >Effect of ribs in HIFU beam path on formation of coagulative necrosis in goat liver
【24h】

Effect of ribs in HIFU beam path on formation of coagulative necrosis in goat liver

机译:肋骨束路径肋骨叶片在山羊肝脏凝固坏死形成上的影响

获取原文

摘要

The motives of the work are to explore the effect of ribs in HIFU beam path on HIFU ablation goat liver. A model-JC Focused Ultrasound Tumor Therapeutic System was used. A 0.75 MHz focused transducer with 150mm aperture and 19.0 mm focal length was used in all experiment. Acoustical power can be adjusted. 30 goats were divided into control group (HIFU beam through rib cage, HIFU alone), experiment group 1(HDFU beam through rib cage, HIFU combined with microbubble) and experiment group 2(Ribs in HIFU beam path were surgically removed, HIFU alone). 20 targeted regions at 5cm away from skin surface were applied for creating necrosis with linear scanning of 15mm length using HIFU in 3 groups. All animals were sacrificed two days later and exposed organs were dissected. After obtaining the maximal section, the volumes of the necrotic regions were measured, then to calculate Energy Efficiency Factor (EEF). Researched results showed that Ribs in HIFU beam path affected the formation of coagulative necrosis and enhanced EEF in control group. HIFU combined with microbubble could enhance the formation of coagulative necrosis and decrease EEF.
机译:该工作的动机是探讨肋骨在Hifu烧蚀山羊肝脏上的HIFU梁路径的影响。使用模型-JC聚焦超声肿瘤治疗系统。所有实验中使用了0.75 MHz的聚焦传感器,孔径为150mm孔径和19.0mm焦距。可以调整声音功率。将30只山羊分为对照组(HIFU梁通过肋骨,单独的HIFU),实验组1(HIFU梁通过肋骨,HIFU与微泡)和实验组2(HIFU梁路径中的肋骨外,单独除去HIFU梁路径) 。距离皮肤表面5厘米的20个靶向区域施用用于产生任何3毫米长度的粗糙扫描,使用3组。两天后牺牲了所有动物,并解剖了暴露的器官。在获得最大部分之后,测量坏死区域的体积,然后计算能效系数(EEF)。研究结果表明,HIFU梁路径中的肋骨影响了对照组凝固坏死和增强eEF的形成。 HIFU与Microbubble相结合可以增强凝固坏死和减少eef的形成。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号