In the first half of 2005, Nederlands Meetinstituut Van Swinden Laboratorium B.V. (NMi VSL) redesigned their facilities for radiation thermometry in a new laboratory building and an opportunity arose to implement new measurement methods. The new facility is used for ITS-90 realization and dissemination in the temperature range from -50 deg C to 3,000 deg C. A study was performed to compare a silver-point realization with a fixed-point blackbody radiator (FP-BBR) to a sodium heat-pipe blackbody radiator (HP-BBR) traceable via a HTSPRT to a contact thermometry silver point. It was found that the fixed-point realization transfer to the sodium heat pipe results in an uncertainty from 0.2 K to 2.4 K for the ITS-90 over the temperature range from 961.78 deg C to 3,000 deg C.
机译:2005年上半年,Nederlands MeetInstituut Van Swinden Laberium B.V.(NMI VSL)重新设计了他们在新的实验室建筑中的辐射温度的设施,并机会实现了新的测量方法。新设施用于其-90的实现,并在-50℃至3,000℃的温度范围内传播。进行研究以比较用固定点黑体散热器(FP-BBR)的银点实现。通过HTSPRT到接触温度银点可追踪的热管黑体散热器(HP-BBR)。发现固定点实现转移到钠热管的转移导致0.2k至2.4k的不确定度,在961.78℃至3,000℃的温度范围内的90 k至2.4k。