Measurement of Normal Spectral Emissivity of Niobium (at 633nm) by Pulse-heating Reflectometric Technique

机译:脉冲加热反射技术测量铌(633nm)的正常光谱发射率

获取原文

摘要

Pulse-heating experiments were performed on niobium strip specimen, taking the specimen from room temperature to the melting point based on rapid resistive self-heating of the specimen. The normal spectral emissivity of the specimen was measured by integrating sphere reflectometry developed at HIT. At the same time, the radiance temperature is measured by high-speed pyrometer from 1100K to the melting point. Details of the measurement technique, measurement apparatus and the calibration technique are described. Results of the normal spectral emissivity of niobium at 633nm from room temperature to its melting point are presented and discussed.
机译:在铌条状试样上进行脉冲加热实验,基于试样的快速电阻自加热,将样品从室温从室温中取出到熔点。通过在击中产生的球体反射测量法测量样品的正常光谱发射率。同时,通过高速高温计从1100K到熔点测量辐射温度。描述了测量技术,测量装置和校准技术的细节。介绍并讨论了从室温下633nm处的铌的正常光谱发射率的结果,并讨论了其熔点。