Implementation of PLTS-2000: He-3 Melting Pressure Temperature Scale

机译:实施PLTS-2000:HE-3熔化压力温度级

获取原文

摘要

The recently adopted PLTS-2000, which extends the temperature scale to 0.9 mK, is defined by the melting pressure of ~3He and various fixed points on the melting curve. Because of the minimum in P(T) at 315 mK, pressures at lower temperatures must be measured with an in situ transducer. In this paper, the equipment necessary for producing melting helium and measuring the pressure are described. This includes the Straty-Adams capacitive pressure transducer, the bridge for measuring capacitance, and the ~3He gas handling system. Procedures for calibrating the pressure transducer and for observing the various fixed points established in PLTS-2000 are presented.
机译:最近采用的PLTS-2000,其将温度级延伸至0.9 mk,由〜3He的熔化压力和熔化曲线上的各种固定点定义。由于315 mk的p(t)的最小值,必须用原位换能器测量较低温度的压力。在本文中,描述了制备熔化氦并测量压力所需的设备。这包括Straty-ADAMS电容式压力传感器,用于测量电容的桥梁,以及〜3HE气体处理系统。提出了校准压力传感器和用于观察在PLTS-2000中建立的各种固定点的程序。