Bilateral ITS-90 Comparison at WC-C Peritectic Fixed Point Between NIM and NPL

机译:双边IT-90在NIM和NPL之间的WC-C包层固定点比较

获取原文

摘要

The WC-C peritectic fixed point, nominal melting and freezing temperature 2747 °C, shows extremely good metrological potential. Elsewhere, we published a prototype scale comparison of the ITS-90 between NPL, NIM and CEM, using high temperature eutectic fixed points (HTFPs) of Co-C (1324 °C), Pt-C (1738 °C), and Re-C (2474 °C). In this paper we present the further results of the bilateral comparison of the ITS-90 at an even higher temperature, 2747 °C, between NIM and NPL using WC-C peritectic fixed points. A NIM single zone high temperature furnace, model Chino IR-80, was modified to extend its temperature to 2800 °C. Then, an NPL researcher, on secondment to NIM, filled two WC-C cells in the modified furnace in a vertical position. The two WC-C cells were then realized in the same furnace, in an horizontal position. Their melting temperatures, defined by the inflection point of the melting curves, were measured by a linear pyrometer, model NIM-PSP. NIM's ITS-90 scale was assigned to the two cells, which were then transported to NPL. The realization of NPL's ITS-90 was then assigned to the two cells by using a model HT9500 Thermogauge furnace to realize the fixed points and a linear pyrometer, model LP3, to determine their temperature. The difference from the mean value of the NIM and NPL ITS-90 values for the WC-C points was derived. This allowed us to compare ITS-90 as realized by the two institutes and to determine the uncertainty in the scale comparison.
机译:WC-C包层固定点,标称熔化和冷冻温度2747°C,显示出极高的计量潜力。在其他地方,我们在NPL,NIM和CEM之间发布了ITS-90的原型比较,使用CO-C(1324°C),PT-C(1738°C)和RE的高温共晶定点(HTFP)和RE -C(2474°C)。在本文中,我们使用WC-C包层固定点在NIM和NPL之间的甚至更高温度下,在甚至更高的温度下,在甚至更高的温度下,在甚至更高的温度下,在甚至更高的温度下,在甚至更高的温度下,使用WC-C包层固定点。 MODIEN SING SINGLO型高温炉,型号奇潮IR-80,被修饰,将其温度延伸至2800°C。然后,NPL研究人员在借调到NIM,在改进的炉中填充两个WC-C细胞在垂直位置。然后将两个WC-C细胞在相同的炉中实现,在水平位置。它们的熔化温度由熔化曲线的拐点定义,通过线性高温计,模型NiM-PSP测量。 Nim的ITS-90刻度被分配给两种细胞,然后将其运输到NPL。然后通过使用型号HT9500热摩托炉将NPL的ITS-90分配给两个单元,以实现固定点和线性高温计,型号LP3,以确定其温度。推导出与WC-C点的NIM和NPL ITS-90值的平均值的差异。这允许我们将其-90比较由两所研究所的实现,并确定规模比较中的不确定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号