首页> 外文期刊>Metrologia: International Journal of Scientific Metrology: = Internationale Zeitschrift fur Wissenschaftliche Metrologie: = Journal International de Metrologie Scientifique >Estimation of the degree of equivalence of the ITS-90 above the silver point between NPL, NIM and CEM using high temperature fixed points
【24h】

Estimation of the degree of equivalence of the ITS-90 above the silver point between NPL, NIM and CEM using high temperature fixed points

机译:使用高温固定点估算ITS-90在NPL,NIM和CEM之间的银点以上的等效程度

获取原文
获取原文并翻译 | 示例
           

摘要

Formal pair-wise temperature differences, a comparison reference value (CRV) and degree of equivalence, between the UK (NPL), Chinese (NIM) and Spanish (CEM) national measurement institutes (NMIs), are calculated from the results of a prototype key comparison of the ITS-90 above the silver point using high temperature fixed points (HTFPs) of Co-C (1324 deg C), Pt-C (1738 deg C) and Re-C (2474 deg C). The local realizations of ITS-90 temperatures were assigned by NPL, NIM and CEM to their individual sets of HTFPs. NIM and CEM then transported their cells to NPL and the temperature was realized by all the cells determined in quasi-identical measurement conditions using a linear pyrometer. From these results the pair-wise temperature differences between the three institutes were established. In addition a non-weighted and a weighted CRV were calculated for the comparison and degree of equivalence estimated. The results show very close agreement of the ITS-90 as realized by the three institutes. It is argued that the ease of the comparison and the low uncertainties attained strongly suggest that HTFPs should be the scale comparison artefacts of choice when comparing primary realizations either of the ITS-90 or thermodynamic temperature.
机译:根据原型结果计算出英国(NPL),中国(NIM)和西班牙(CEM)国家测量机构(NMI)之间的正式成对温差,比较参考值(CRV)和等效程度使用Co-C(1324摄氏度),Pt-C(1738摄氏度)和Re-C(2474摄氏度)的高温固定点(HTFP)对ITS-90高于银点的关键比较。 NPL,NIM和CEM将ITS-90温度的本地实现分配给它们各自的HTFP。然后,NIM和CEM将其电池运输到NPL,并使用线性高温计在准相同的测量条件下确定的所有电池实现温度。根据这些结果,确定了三个研究所之间的成对温差。此外,还计算了非加权CRV和加权CRV,以进行比较并估算出等效程度。结果表明,这三个研究所实现了ITS-90的高度一致。有人认为,比较的容易性和所获得的低不确定性强烈表明,在比较ITS-90或热力学温度的主要实现方法时,HTFP应该是规模比较伪像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号