首页> 外文会议>International Workshop on Digital Mammography >Performance of the New Double Layer Amorphous Selenium Detector for Digital Mammography Compared to the FDA Approved CR System for Digital Mammography
【24h】

Performance of the New Double Layer Amorphous Selenium Detector for Digital Mammography Compared to the FDA Approved CR System for Digital Mammography

机译:用于数字乳房X线摄影的新型双层无定形硒检测器的性能与FDA批准的数字乳房X线摄影

获取原文

摘要

The performance of a new double layer amorphous Selenium detector for digital mammography is compared to the FDA approved CR system for digital mammography in a two step study. A study comparing radiation between both systems is first done to obtain the best settings for a clinical study. In a second step the results in terms of quality are evaluated by three readers comparing the final images. A minimal reduction of dose (20%) is obtained and a better definition of glandular structures is demonstrated with the new device.
机译:将用于数字乳房X线摄影的新的双层无定形硒检测器的性能与两步研究中的FDA认可的CR系统进行比较。首先进行比较两个系统之间的辐射以获得临床研究的最佳设置。在第二步中,通过比较最终图像的三个读者评估质量方面的结果。获得剂量(20%)的最小减少,并用新装置对腺体结构进行更好的定义。