Hyper-sensitivity to nickel and molybdenum ions may be a correlatory factor to in-stent restenosis after implantation of conventional coronary stents. The "implantation" of negative carbon ions into the stainless steel of the Arthos-Inert? stent reduces ion diffusion into the vessel wall. The present registry aims to report the clinical findings in 121 patients after the implantation of this stent. Angiographic follow-up was available for 22 patients, whereof 10 showed in-stent restenosis (8%). Clinical restenosis, including patients with angiographic restenosis or angina, was 13%. Only 8 patients underwent TLR, whereof 2 had developed myocardial infarction, and 2 cardiac deaths were reported. Thus, overall MACE rate was 8%. This registry showed good clinical results after implantation of the Arthos-Inert? stent in a large collective.
机译:对镍和钼离子的超敏感性可能是植入常规冠状支架后支架内再狭窄的相关因素。负碳离子的“植入”到惰性液体的不锈钢中?支架将离子扩散降低到血管壁中。本书旨在在植入这座支架后在121名患者中报告临床发现。 22名患者可获得血管造影随访,其中10例显示出含有支架再狭窄(8%)。临床再狭窄,包括血管造影或心绞痛患者的患者为13%。只有8名患者接受了TLR,其中2例开发了心肌梗死,报告了2例心脏死亡。因此,总体坐标率为8%。该注册表在植入氨基酚后显示出良好的临床结果?支架在一个大集体中。