首页> 外文会议>IEEE International Conference on Sensors >Micro Analysis System for pO_(2), pCO_(2), and pH Constructed with Stacked Modules
【24h】

Micro Analysis System for pO_(2), pCO_(2), and pH Constructed with Stacked Modules

机译:PO_(2),PCO_(2)的微分析系统,并用堆叠模块构建的pH

获取原文

摘要

A miniaturized analysis system for on-line monitoring of pO_(2), pCO_(2), pH was constructed by stacking a reference electrode module, a sensing module, and a sampling module. The liquid-junction reference electrode module used durable thin-film Ag/AgCl electrodes to use an internal solution saturated with KCl. A Clark-type pO_(2) sensor, a Severinghaus-type pCO_(2) sensor, and an indicator electrode for a pH sensor were integrated on a sensing module. The sampling module used reversible growth or shrinkage of a hydrogen bubble. The volume change of the gas was converted to the movement of the solution in a flow channel by a silicone rubber diaphragm. The responses of the three sensors were checked by introducing external standard solutions into the flow channel by the sampling module. Clear responses and linear calibration curves were observed for the respective sensors.
机译:通过堆叠参考电极模块,传感模块和采样模块,构建用于PO_(2),PCO_(2),PH构造的小型化分析系统。液 - 结参考电极模块使用耐用的薄膜Ag / AgCl电极,使用与KCl饱和的内溶液。克拉克型PO_(2)传感器,Severinghaus型PCO_(2)传感器和PH传感器的指示器电极集成在传感模块上。采样模块使用氢气泡的可逆生长或收缩。通过硅橡胶膜片将气体的体积变化转化为溶液在流动通道中的运动。通过采样模块将外部标准解决方案引入流量通道来检查三个传感器的响应。对于各个传感器,观察到清除响应和线性校准曲线。