首页> 外文会议>Air Waste Management Association's Annual conference Exhibition >ETV Test of Opsis Open-Path UV-DOAS Monitor for Measurements of Nitric Oxide (NO), benzene (C6H6) and Ammonia (NH3)
【24h】

ETV Test of Opsis Open-Path UV-DOAS Monitor for Measurements of Nitric Oxide (NO), benzene (C6H6) and Ammonia (NH3)

机译:OPSIS开放式UV-DOAS监测器的ETV测试,用于测量一氧化氮(NO),苯(C6H6)和氨(NH3)

获取原文

摘要

The Opsis UV-DOAS monitor was ETV verified in April, 2000, for measurements of nitric oxide (NO), benzene (C6H6) and ammonia (NH3). The test included measurements of the target gases at different concentrations, path lengths, integration times, and source intensities to assess the minimum detection limit (MDL), source strength linearity, concentration linearity, accuracy, precision, and sensitivity to atmospheric interferences. The measured minimum detection limits at the 800 ft. (250 meter) test distance for nitric oxide, benzene and ammonia were 0.9, 0.5 and 2.8 ppb respectively. This paper provides a description of the UV-DOAS system configuration, the test procedure and a summary of the test results.
机译:OPSIS UV-DOAS监测仪于2000年4月验证了ETV,用于测量一氧化氮(NO),苯(C6H6)和氨(NH3)。该测试包括在不同浓度,路径长度,集成时间和源强度下的目标气体的测量,以评估最小检测限(MDL),源强度线性,浓度线性,精度,精度和对大气干扰的敏感性。测量的800英尺的最小检测限度分别为0.9,0.5和2.8ppb的800英尺。(250米)的测试距离。本文提供了对UV-DOAS系统配置,测试程序和测试结果摘要的描述。