REALIZATION OF THE ITS-90 AT THE NRLM USING ~(4)He VAPOR-PRESSURE THERMOMETRY

机译:使用〜(4)HE蒸汽压力测温法在NRLM处实现其-90

获取原文

摘要

The ~(4)He vapor-pressure temperature scale is realized in the range of 2.2 K to 4.3 K as part of the realization of the International Temperature Scale of 1990 at the National Research Laboratory of Metrology (NRLM), Japan. The vapor pressure is measured using an absolute pressure gauge set at room temperature. The sensing element of the gauge is a silicon resonant sensor attached to a diaphragm. The temperature distribution along the pressure-sensing tube is measured using five thin-film Rh-Fe resistance thermometers. Three capsule-type Rh-Fe resistance thermometers are calibrated against the scale and the measurement uncertainties in the realization and the calibration are estimated. The combined standard uncertainty of the fitted equation is less than 0.1 mK from 2.25 K to 4.25 K for each thermometer. The present realization is compared with the ~(4)He vapor-pressure scale realized in 1976 at the NRLM by calibrating the same Rh-Fe thermometers against both scales. The difference between the two scales is within 0.2 mK from 2.25 K to 4.25 K and is less than the combined standard uncertainty of measurement.
机译:〜(4)他气压温度秤在2.2 k到4.3 k的范围内实现,作为1990年国际测量研究实验室(NRLM),日本的1990年国际温度等级的一部分。使用在室温下设定的绝对压力计测量蒸汽压力。表的传感元件是附接到隔膜的硅谐振传感器。使用五薄膜RH-FE电阻温度计测量沿压力传感管的温度分布。三个胶囊型RH-FE电阻温度计抵抗尺度校准,估计了实现和校准中的测量不确定性。对于每个温度计,安装方程的合并标准不确定性小于2.25k至4.25 k的0.1 mk。将本发明与1976年在NRLM实现的〜(4)的蒸气压秤进行比较,通过校准相同的RH-FE温度计抵抗两个尺度。两种尺度之间的差异在2.25 k至4.25 k的0.2 mk内,并且小于测量的组合标准不确定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号