Developing a Device for Monitoring O_2 Saturation in Blood

机译:开发用于监测血液中O_2饱和的装置

获取原文

摘要

We present a single-channel device for monitoring oxygen saturation in blood using near infrared light (NIR). This device measures the differential absorption of both oxi-hemoglobin (HbO_2) and deoxi-hemoglobin (HHb) concentrations using wavelengths within the biophotonic window (700 nm - 1100 nm). Our device works with two wavelengths:the first source (λ_1 = 632 nm) operates in continuous wave (CW) and the second (λ_2 = 940 nm) operates in pulsed mode. The pulsed signal at λ_2 operates in a 20% fill factor. The CW signal provides information related to the arterial pulse, whereas both CW and pulsed signal provide information about blood oxygenation. Light emitted from the sources travels through the tissue and are collected on the detector by transmission. Measurements of blood oxygenation measured in transmission on fingers tips are presented which are comparable to commercial devices. The values obtained are above 90.0% blood oxygenation in healthy tests subjects with 0.1% accuracy in the measurements.
机译:我们介绍了一种用于在近红外光(NIR)的血液中监测血液中的氧饱和的单通道装置。该装置测量使用生物光学窗口(700nm-1100nm)内的波长的氧气血红蛋白(HbO_2)和脱氧血红蛋白(HHB)浓度的差异吸收。我们的设备适用于两个波长:第一源(λ_1= 632nm)在连续波(CW)中操作,第二(λ_2= 940nm)以脉冲模式操作。 λ_2的脉冲信号以20%的填充因子操作。 CW信号提供与动脉脉冲相关的信息,而CW和脉冲信号均提供有关血氧的信息。从源发出的光通过组织行进,并通过透射在探测器上收集。提出了在手指尖端上测量的血氧测量的测量,其与商业设备相当。所得值高于90.0%的血氧在测量中具有0.1%精度的受试者的血氧化。