In this work the new data acquisition system (DAQ) for the microPET of the SIBI project is presented. To increase the microPET sensitivity, the inclusion of more detection modules is required, which in turn needs a more sophisticated and compact signal processing system. The new DAQ is based on programmable integrated circuits (FPGAs) and is composed of (i) an 8-input triggering board with individual channel adjusting capabilities, which can process signals from 8 detector modules working in coincidence mode and (ii) two 10-channel digitising boards with 12-bit resolution. The digitised signals are transmitted to a PC through two Ethernet ports in each board. With the new boards the maximum singles counting rate is of the order of 350 kHz, with a dead time of 2.8 ps. Individual crystal maps of two detectors for image corrections have been obtained, with peak-to-valley ratios of 5:1. The new FPGA boards will allow the introduction of more detection modules with relatively simple electronics arrangement.
机译:在这项工作中,介绍了SIBI项目的Micropet的新数据采集系统(DAQ)。为了提高微移泌灵敏度,需要包含更多的检测模块,这又需要更复杂和紧凑的信号处理系统。新DAQ基于可编程集成电路(FPGA),由(i)一个带有各个频道调节功能的8输入触发板组成,其可以从巧合模式和(ii)中工作的8个检测器模块的信号处理两个10-具有12位分辨率的通道数字化板。数字化信号通过每个板中的两个以太网端口传输到PC。通过新的电路板,最大单打数计数率为350 kHz,死区时间为2.8 ps。已经获得了两个探测器的各个晶体图,其中峰值比率为5:1。新的FPGA板将允许引入具有相对简单的电子设备的更多检测模块。