首页> 外文会议>Annual International Conference on Micro Electro Mechanical Systems >UNIVERSAL CONCEPT FOR FABRICATING ARBITRARY SHAPED μIPMC TRANSDUCERS AND ITS APPLICATION ON DEVELOPING ACCURATELY CONTROLLED SURGICAL DEVICES
【24h】

UNIVERSAL CONCEPT FOR FABRICATING ARBITRARY SHAPED μIPMC TRANSDUCERS AND ITS APPLICATION ON DEVELOPING ACCURATELY CONTROLLED SURGICAL DEVICES

机译:用于制造任意形状的μIPMC传感器的通用概念及其在开发精确控制的外科手术装置上的应用

获取原文

摘要

This paper presents a novel micromachined method for fabricating micro-sized Ion Polymer Metal Composite (IPMC) devices with arbitrary shapes and demonstrates the technique to develop tiny accurately controlled surgical devices. This technology is beyond current state-of-the-art approaches in manufacturing large-scale IPMC devices, and provides a universal solution to construct micro-size-featured IPMC transducers with fine contours and no short-circuits. To demonstrate the usefulness of the method, hand-shaped grips (Fig. 1) for laparoscopic operation are designed, fabricated and tested. The fabricated device with the dimension of 6mm × 6mm × 80 μm shows that it could reach a displacement of 300μm at the location 1mm away from the anchor when a 0.5Hz square wave of 12 Volts is applied, and a maximum instantaneous force of 5mN output when it is driven with a 0.5Hz square wave of 6 Volts.
机译:本文介绍了一种用于制造具有任意形状的微型离子聚合物金属复合物(IPMC)器件的新型微机械方法,并演示了开发微精确控制的外科手术装置的技术。该技术超出了现有的最新方法,在制造大型IPMC设备方面,提供了一种通用的解决方案,用于构建具有精细轮廓和没有短路的微尺寸特色IPMC传感器。为了证明该方法的有用性,设计,制造和测试了用于腹腔镜操作的手形夹具(图1)。具有6mm×6mm×80μm的尺寸的制造装置表明,当施加0.5Hz平方波的0.5Hz平方波时,它可以在远离锚点的位置1mm的位置达到300μm的位移,以及5Mn输出的最大瞬时力当它被驱动时的0.5Hz平方波6伏。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号