首页> 外文会议>European Conference on Solid-State Transducers >Application of ionic liquid to the construction of Cu(II) ion-selective electrode with solid contact
【24h】

Application of ionic liquid to the construction of Cu(II) ion-selective electrode with solid contact

机译:离子液体在Cu(II)离子选择电极与固体接触的施工

获取原文

摘要

Potentiometric properties of copper ion-selective electrodes (ISEs) with solid contact based on polymeric membrane were investigated. For the electrode construction ionic liquids (ILs), alkylmethylimidazolium chlorides are used as transducer media. The addition of ionic liquid to the membrane phase was found to improve the analytical parameters of the studied electrodes. The best results were obtained for electrode having membrane doped with 1-ethyl-3 -methyl imidazolium chloride. The electrode shows a Nernstian response for copper ions over a wide concentration range (1x10~(-7)-1x10~(-1) mol L~(-1)) and the slope of 28.9 mV/decade. The limit of detection is 3.2x10~(-8) mol L~(-1). It has a fast response time of 5-10 s and can be used for more than 4 months without any divergence in potential. The proposed sensor is not pH sensitive in the range 2.5-6.0 and shows a very good discriminating ability towards Cu~(2+) ion in comparison with some alkali, alkaline earth, transition and heavy metal ions.
机译:研究了基于聚合物膜的固体触点的铜离子选择性电极(ISE)的电位性能。对于电极结构离子液体(ILS),烷基甲基甲基咪唑氯化物用作换能器介质。发现在膜相中加入离子液体以改善所研究的电极的分析参数。获得具有掺杂有1-乙基-3-甲基甲基咪唑鎓氯化锆的电极的最佳结果。电极显示铜离子在宽浓度范围内的Nernstian响应(1×10〜(-7)-1x10〜(-1)mol l〜(-1)),斜率为28.9 mV /十年。检测限为3.2x10〜(-8)mol l〜(-1)。它具有5-10秒的快速响应时间,可用于超过4个月而不会出现潜力。所提出的传感器在2.5-6.0的范围内不pH敏感,与一些碱,碱土,过渡和重金属离子相比,朝向Cu〜(2+)离子的极好区分能力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号