首页> 外文会议>Annual scientific meeting of computers in cardiology >Influence of low sampling rate on heart rate variability analysis based on non-linear dynamics
【24h】

Influence of low sampling rate on heart rate variability analysis based on non-linear dynamics

机译:低采样率对非线性动力学心率变异分析的影响

获取原文

摘要

The aim of this study was to examine how sampling rates below 1000 Hz influence the precision of heart rate variability analysis. Simulated tachograms as well as tachograms from healthy persons were investigated. With parameters from time, frequency domain and non-linear dynamics these tachograms were classified. The results were compared with theoretical assumptions and special error measures were derived. New developed measures from non-linear dynamics showed an insensitive behavior to a reduced sampling frequency down to 128 Hz. The major results of this study lead to the conclusion to use sampling rates only above 120 Hz which implies a total error in the range of approximately 1.8/spl plusmn/1.5%. However only the use of sampling rates above 200 Hz provides nearly error free results.
机译:本研究的目的是检查低于1000Hz的采样率如何影响心率可变性分析的精度。调查了模拟的快速图以及来自健康人的快速图。对于时间,频域和非线性动力学的参数,这些Tachogks被分类。结果与理论假设进行了比较,衍生出特殊的误差措施。来自非线性动力学的新开发措施向降低的采样频率降低至128Hz,对不敏感的行为显示出不敏感的行为。本研究的主要结果导致仅使用上高于120 Hz的采样率,这意味着大约1.8 / SPL PLUSMN / 1.5%的总误差。然而,只有200 Hz的采样率的使用提供了几乎无错误的结果。