首页> 外文会议>International conference on computer applications in biotechnology >THE ON-LINE COMPARISON OF A MEASURED VS CALCULATED OXYGEN MASS TRANSFER COEFFICIENT - A USEFUL PARAMETER FOR FAULT DIAGNOSIS
【24h】

THE ON-LINE COMPARISON OF A MEASURED VS CALCULATED OXYGEN MASS TRANSFER COEFFICIENT - A USEFUL PARAMETER FOR FAULT DIAGNOSIS

机译:测量的VS的在线比较计算氧气传质系数 - 一种有用的故障诊断参数

获取原文

摘要

The equations for the derivation of a measured and calculated oxygen transfer coefficient are briefly described. It is then shown how an on-line comparison of these two values can be used to obtain process information for run optimisation, instrument fault detection, overdosing of reagents such as antifoam and irregular changes of broth rheology. A single example showing the unambiguous detection of fouling on a dissolved oxygen probe is finally presented and discussed.
机译:简要描述了用于测量和计算的氧传输系数的推导的等式。然后示出了这两个值的在线比较如何用于获得运行优化,仪器故障检测,过量试剂的过程信息,例如消泡剂和肉汤流变的不规则变化。最后介绍并讨论了显示在溶解的氧气探针上污垢污垢的明确检测的单一示例。