External Ultrasonic Valvuloplasty for the Treatment of Superficial Venous Insufficiency: A feasibility study

机译:外部超声波瓣膜成形术治疗浅表静脉功能不全:可行性研究

获取原文

摘要

Deformations on the valvular tissues of the saphenous vein can cause an abnormal reflux of blood. This reflux is the main cause of Superficial Venous Insufficiency (SVI) and varicose veins. Since High Intensity Focused Ultrasound (HIFU) nduces a strong rise in the temperature of biological tissue, and collagen in veins deforms when exposed to a high temperature, we conducted a study to demonstrate the feasibility of using HIFU to induce shrinkage of the collagen of the saphenous vein near the valvular tissue. This shrinkage should restore normal function of the valve. An experimental protocol was carried out where several in vitro segments of human saphenous vein were sonicated with HIFU. In order to better recreate in vivo conditions, the segment was inserted in porcine muscle. The segment of vein and the porcine muscle sample were placed into a PVG cylinder. After being degassed, the sample was inserted in a recipient filled with degassed-s aline solution and sonicated with a real-time imaging HIFU probe. The probe has a focal length of 45 mm, a diameter of 52.5 mm and operates at 3 MHz. Ultrasonic maging, obtained by an 8-MHz 128-element linear array placed n the center of the HIFU probe, was used to spot the focal zone on the wall of the vein. Individual sonications of the vein wall were performed for acoustic power values ranging between 8.75 and 35 W at a constant sonication duration of 5-s. Different tonication durations ranging between 3 and 7-s at constant power were also tested. Results showed that shrinkage of the vein wall was observed by echographic and macroscopic analysis for the cases of high-power sonication. Histological analysis showed that collagen of the vein wall was indeed altered for all cases of sonications but no signs of perforation were found. However, a macroscopic ablation was observed in the surrounding muscle. The observed shrinkage of the vein wall suggests that dysfunctional valves of the saphenous vein could be treated with HIFU since a significant shrinkage of collagen could restore competence of the abnormal valve.
机译:在隐静脉的瓣膜组织的变形可能会导致血液异常回流。此回流是浅表静脉功能不全(SVI)及静脉曲张的主要原因。由于高强度聚焦当暴露于高温下超声(HIFU)nduces在生物组织的温度的强上升,胶原在静脉变形,我们进行了研究,以证明使用HIFU诱导的胶原收缩的可行性附近的瓣膜组织隐静脉。这种收缩应该恢复阀门的正常功能。的实验方案进行,其中在人隐静脉的体外段几个用HIFU超声处理。为了在体内条件更好重新创建,段插入在猪肌肉。静脉和肌肉猪样品的段放入一个汽缸PVG。脱气后,将样品插入填充有脱气-S艾琳溶液并超声处理以实时成像HIFU探头的收件人。探针具有45mm的焦距,直径52.5毫米,3 MHz运行。超声波maging,由8 MHz的128元件线性阵列置于n中的HIFU探头的中心获得的,使用至斑点聚焦区上的静脉壁。静脉壁的个别超声处理的声学功率值8.75和W之间的范围35在5-s的恒定超声处理持续时间进行。不同持续时间tonication范围3和7之间-s的恒定功率也进行了测试。结果表明,通过对大功率超声处理的情况下,回波描记和宏观分析中观察到静脉壁的收缩那个。组织学分析显示静脉壁的胶原蛋白确实是改变了声处理的所有案件,但没有发现穿孔的迹象。但是,在周围的肌肉中观察到肉眼可见的烧蚀。静脉壁的收缩观察表明,隐静脉的功能失调的阀可以与HIFU进行处理,因为胶原的显著收缩能够恢复异常阀的能力。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号