Respiratory-gated micro-CT using a carbon nanotube based micro-focus field emission x-ray source

机译:使用基于碳纳米管的微焦点场发射x射线源进行的呼吸门控微型CT

获取原文

摘要

A prototype physiologically gated micro-computed tomography (micro-CT) system based on a field emission micro-focus x-ray source has been developed for in vivo imaging of small animal models. The novel x-ray source can generate radiation with a programmable waveform that can be readily synchronized and gated with non-periodic physiological signals. The system performance is evaluated using phantoms and sacrificed and anesthetized mouse models. Prospective respiratory-gated CT images of anesthetized free-breathing mice are collected using this scanner at 100 msec temporal resolution and 101p/mm of 10% system MTF.
机译:基于场发射微焦点x射线源的原型生理门控微计算机断层扫描(micro-CT)系统已经开发出来,可用于小动物模型的体内成像。新颖的X射线源可以产生具有可编程波形的辐射,该辐射可以很容易地与非周期性生理信号同步并进行门控。使用体模以及牺牲和麻醉的小鼠模型评估系统性能。使用该扫描仪以100毫秒的时间分辨率和101p / mm的10%系统MTF收集麻醉的自由呼吸小鼠的前瞻性呼吸门控CT图像。