首页> 外文会议>IEEE International Conference on Cyborg and Bionic Systems >Optical Measurement of Deformation Distribution on Retinal Model for Vitreoretinal Surgery Training
【24h】

Optical Measurement of Deformation Distribution on Retinal Model for Vitreoretinal Surgery Training

机译:玻璃体视网膜手术训练的视网膜模型上变形分布的光学测量

获取原文

摘要

Training of vitreoretinal surgeries such as peeling of inner limiting membrane (ILM) requires elaborate retina model having contact measurement function since human retina is the delicate tissue. Fabrication of the retina model having measurement function without impairing the physical properties of the human retina is challenging issue. This paper reports the development of a retina model for optical measurement of deformation distribution using digital image correlation (DIC). The retinal model consists of the retina part and the sclera part. Both parts reproduce Young's modulus of human retina and sclera, respectively. The deformation distribution on the surface of retinal model was measured by the non-contact measurement using DIC. The Young's modulus of the retina part and sclera part were 30 kPa and 1.9 MPa, respectively. We demonstrated the optical measurement of deformation distribution on the surface of retinal model during the pushing with ILM forceps.
机译:玻璃体视网膜手术的训练,例如内膜限制膜(ILM)的剥离,需要复杂的具有接触测量功能的视网膜模型,因为人的视网膜是脆弱的组织。具有测量功能的视网膜模型的制造而不损害人类视网膜的物理特性是具有挑战性的问题。本文报道了使用数字图像相关(DIC)光学测量变形分布的视网膜模型的发展。视网膜模型由视网膜部分和巩膜部分组成。这两部分分别再现人视网膜和巩膜的杨氏模量。通过使用DIC的非接触式测量来测量视网膜模型表面上的变形分布。视网膜部分和巩膜部分的杨氏模量分别为30 kPa和1.9 MPa。我们展示了用ILM钳推动过程中视网膜模型表面变形分布的光学测量。