首页> 外文会议>International Conference on Medical Imaging and Computer-Aided Diagnosis >Optical Micro-scanning Reconstruction Technique for a Thermal Microscope Imaging System
【24h】

Optical Micro-scanning Reconstruction Technique for a Thermal Microscope Imaging System

机译:热显微镜成像系统的光学微扫描重建技术

获取原文

摘要

Thermal microscope imaging system can achieve the detection of sub-tle targets by temperature changes, which has extensive utility value in some respects that need detailed thermal analysis. Due to the mechanical preparation of the system, working vibration, external environment and other factors of inter-ference, which will affect the precision of the micro-scanning system, Causing the micro-displacement of the four images acquired by the system deviate from the standard position. Therefore affecting the quality of subsequent image recon-struction. To minimize the impact of these factors, a micro-scanning reconstruction technique based on local gradient interpolation was proposed. The simulation and experiment demonstrate that the method can decrease the system micro-scanning error and enhance image reconstruction quality.
机译:热显微镜成像系统可以通过温度变化实现对细小目标的检测,在某些方面需要详细的热分析,具有广泛的实用价值。由于系统的机械准备,工作振动,外部环境和其他干扰因素,这些都会影响微扫描系统的精度,从而导致系统获取的四个图像的微位移偏离了扫描系统。标准位置。因此影响了后续图像重建的质量。为了最小化这些因素的影响,提出了一种基于局部梯度插值的微扫描重建技术。仿真和实验表明,该方法可以减少系统的微扫描误差,提高图像重建质量。