An Amplifier-Less Calibration-Free SAR ADC Achieving >100dB SNDR for Multi-Channel ECG Acquisition with 667mVpp Linear Input Range

机译:线性输入范围为667mV pp 的多通道ECG采集,无放大器免校准SAR ADC达到100dB SNDR

获取原文

摘要

This work presents a time-multiplexing SAR ADC to support up to 5-lead ECG monitoring with >100dB SNDR per readout channel. Its noise and linearity performance are enhanced by a combination of dual-reference architecture and mismatch error shaping (MES) technique without using amplifiers or calibration, resulting in >106dB SFDR and 109.4dB DR within 250Hz bandwidth (FoMS,DR=178.9dB). The ECG analog front-end (AFE), including 3 DC-coupled instrumentation amplifiers (IAs) and 1 ADC, occupies only 0.48mm2 in 55nm CMOS. Each ECG channel achieves 1μVrms (0.5-250Hz) input-referred noise at a low IA gain of 6V/V with a 667mVpp-diff linear input range.
机译:这项工作提出了一个时分多路SAR ADC,以每个读出通道支持大于100dB SNDR的5导联ECG监视。通过结合使用双参考架构和失配误差整形(MES)技术而无需使用放大器或校准,可以增强其噪声和线性性能,从而在250Hz带宽(FoM)内提供> 106dB的SFDR和109.4dB的DR S,DR = 178.9dB)。 ECG模拟前端(AFE)包括3个直流耦合仪表放大器(IAs)和1个ADC,仅占0.48mm 2 在55nm CMOS中每个ECG通道均达到1μV rms 在667mV的6V / V的低IA增益下(0.5-250Hz)输入参考噪声 pp-diff 线性输入范围。