首页> 外文会议>SPIE Photonics Europe Conference >Automation of Spectroellipsometric Measurements Within Range of 1-4,9 eV by Beattie-Conn method
【24h】

Automation of Spectroellipsometric Measurements Within Range of 1-4,9 eV by Beattie-Conn method

机译:Beattie-Conn方法在1-4,9 eV范围内自动进行椭偏光度法测量

获取原文

摘要

The spectroellipsometer based on DMR-4 spectrometer was automated. An electronic system and software for the automation of equipment was designed and built. The system was calibrated and for the purpose of verification of optical constants of metals samples such as gold, silver and copper were measured. The device automatically measures metal surfaces by Beattie-Conn method. It was measured values of optical constants of metal and compared them with table data. Graphs were built by experimental and table data, they demonstrate the results of measurements of metal samples coincided with theoretical data. Optical constants of surface based on heterostructures Cr_(1,5nm)Au_(47nm)HfO_(2_7nm) were measured by Beattie-Conn method too.
机译:基于DMR-4光谱仪的光谱仪是自动化的。设计并制造了用于设备自动化的电子系统和软件。对该系统进行了校准,并且为了验证金属样品(如金,银和铜)的光学常数,对其进行了测量。该设备通过Beattie-Conn方法自动测量金属表面。测量金属的光学常数值,并将其与表数据进行比较。图形是通过实验和表格数据构建的,它们证明了金属样品的测量结果与理论数据相吻合。也通过Beattie-Conn方法测量了基于异质结构Cr_(1.5nm)Au_(47nm)HfO_(2_7nm)的表面的光学常数。