Comparison of effects of dose on image quality in digital breast tomosynthesis systems across multiple vendors

机译:多家供应商的数字乳房断层合成系统中剂量对图像质量的影响比较

获取原文

摘要

In traditional radiography and computed tomography (CT), contrast is an important measure of image quality that, in theory, does not vary with dose. While increasing dose may increase the overall contrast-to-noise ratio (CNR), the contrast in an image should be primarily dependent on variation in tissue density and attenuation. We investigated the behavior of all three currently FDA-approved vendors' 3D DBT systems (Siemens, Hologic, and General Electric (GE)) using the Computerized Imaging Reference Systems (CIRS) Model 011A Breast Phantom and found that for both Siemens and Hologic systems, contrast increased with dose across multiple repeated trials. For these two systems, experimental CNR also appeared to increase above the expected CNR, which suggests that these systems seem to have introduced post-processing by manipulation of contrast, and thus DBT data cannot be used to reliably quantify tissue characteristics. Additional experimentation with both 2D mammography and 3D DBT systems from GE in addition to the previously mentioned vendors, however, suggested that this relationship is not true for all systems. An initial comparison of contrast vs. dose showed no relationship between contrast and dose for 2D mammography, with the contrast remaining relatively constant in the dose range of 33% of the automatic exposure control setting (AEC) to 300% AEC for all three vendors. The GE DBT system also did not exhibit increased contrast with increased dose, suggesting that the behavior of 3D DBT systems is vendor-specific.
机译:在传统的放射线摄影和计算机断层扫描(CT)中,对比度是图像质量的重要指标,理论上,它不会随剂量而变化。虽然增加剂量可能会增加总体对比度噪声比(CNR),但图像中的对比度应主要取决于组织密度和衰减的变化。我们使用计算机成像参考系统(CIRS)011A型乳房假体对目前三个获得FDA批准的供应商的3D DBT系统(西门子,Hologic和通用电气(GE))的行为进行了调查,结果发现西门子和Hologic系统都具有这种行为,在多次重复试验中,对比随着剂量的增加而增加。对于这两个系统,实验性CNR似乎也超过了预期的CNR,这表明这些系统似乎是通过操纵对比度引入了后处理的,因此DBT数据不能用于可靠地量化组织特征。除了前面提到的供应商之外,GE还对GE的2D乳腺摄影和3D DBT系统进行了额外的实验,这表明这种关系并不适用于所有系统。造影剂与剂量的初步比较显示,对于2D乳腺摄影,造影剂与剂量之间没有关系,在所有三个供应商的自动曝光控制设置(AEC)的33%到300%AEC的剂量范围内,对比度保持相对恒定。 GE DBT系统也不会随剂量增加而显示出增强的对比度,这表明3D DBT系统的行为是特定于供应商的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号