首页> 外文会议>Conference on Physics of Medical Imaging >Effects of dead time on quantitative dual-energy imaging using a position-sensisitve spectroscopic detector
【24h】

Effects of dead time on quantitative dual-energy imaging using a position-sensisitve spectroscopic detector

机译:死区时间对定量双能成像的影响使用位置激敏光谱探测器

获取原文

摘要

Dual energy imaging (DE) is a potential alternative to conventional mammography for patients with dense breasts. It requires intravenous injection of contrast agent (CA) and subsequent acquisition of images at two different energies. Each pixel is seen as a vector and is projected onto a two-material basis, e.g. water, CA, to form separate water-equivalent and CA-equivalent images. On conventional detectors, this requires two separate exposures. Spectroscopic detectors allow multiple images from a single exposure by integrating appropriate energy bands. This work investigates the effects of high count rates on quantitative DE imaging using a CdTe spectroscopic detector. Because of its small pixel size (250 μm), a limitation of the detector is charge sharing between pixels, which must be corrected to avoid degradation of the detected spectrum. However, as charge sharing is identified by neighbouring pixels registering a count in a given readout frame, an effective maximum count rate (EMR) is imposed, above which linearity between incident and detected counts is lost. A simulation was used to model detector response of a test object composed of water and iodine, with different EMRs and incident count rates. Using a known iodine thickness of 0.03 cm, and an EMR of 10~3 s~(-1), the reconstructed thickness of iodine was found to be 97%, 74% and 24% of the true value for incident count rates of 100, 1000 and 10000 photons/pixel/s respectively. The simulation was validated by imaging a water-equivalent test phantom containing iodinated CA at different X-ray currents, to determine the optimum beam conditions.
机译:双能量成像(DE)是常规X线摄影的潜在替代患者的致密乳房。它要求的造影剂(CA),并在两个不同的能量图像的后续获取的静脉内注射。每个像素被看作是一个向量和被投影到一个两种材料的基础上,例如水,CA,以形成单独的水等价和CA-等效的图像。在常规的探测器,这需要两个单独的曝光。分光检测器允许从单个曝光的多个图像通过集成适当的能量带。这项工作调查的使用碲化镉光谱检测定量DE成像高计数率的影响。由于它的小像素尺寸(250微米)的,检测器的一个限制是像素,这必须予以纠正,以避免所检测的光谱的降解之间的电荷共享。然而,由于电荷共享被邻近在给定的读出帧登记计数像素识别,有效最大计数率(EMR)被施加,高于该事件和检测到的计数之间的线性丢失。仿真中使用的水和碘组成,具有不同的电子病历和入射计数率的测试对象的模型检测器响应。使用碘厚度的0.03厘米已知的,并且10〜3秒〜(-1)的EMR,碘的重建的厚度被发现是97%,74%和100入射计数率的真值的24% ,1000个10000光子/像素/分别秒。模拟是由成像含有碘化CA在不同的X射线电流的水当量测试体模,以确定最佳的波束条件验证。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号