首页> 外文会议>IEEE International Ultrasonics Symposium >Effect of boundary constraints and hydrostatic pressure on the vaporization threshold of low boiling-point phase-change contrast agents
【24h】

Effect of boundary constraints and hydrostatic pressure on the vaporization threshold of low boiling-point phase-change contrast agents

机译:边界约束和静水压力对低沸点相变造影剂汽化阈值的影响

获取原文

摘要

Phase-change contrast agents (PCCAs) can be vaporized using ultrasound for therapeutic and diagnostic applications. It has been previously reported that the in vivo vaporization threshold is higher than in vitro. Here, we explore the role of boundary constraints and increased hydrostatic pressure on the vaporization threshold of PCCAs. Vaporization was performed in tubes of different inner diameters and a chamber under different hydrostatic pressures. Additionally, the vaporization threshold was found in the kidneys of rats. The results show that the vaporization threshold increases with a decreasing tube diameter, and that hydrostatic pressure has no effect of the pressure required for vaporization. The in vivo threshold that was found was higher than the in vitro experiments, but the effect of boundary constraints does not account for all the difference between in vivo and in vitro.
机译:相变造影剂(PCCA)可以使用超声波进行汽化以用于治疗和诊断应用。先前已经报道了体内汽化阈值高于体外。在这里,我们探讨了边界约束和静水压力增加对PCCAs汽化阈值的作用。在不同的内径的管和在不同的静水压力下的腔室中进行汽化。另外,在大鼠的肾脏中发现了汽化阈值。结果表明,汽化阈值随管直径的减小而增加,并且静水压力对汽化所需的压力没有影响。发现的体内阈值高于体外实验,但是边界限制的影响并不能解释体内和体外之间的所有差异。