From the data presented it is clear that issues encountered in the field for HC1 response testing should first look at the materials used, the flow rate of the test gas and the regulator as culprits in the retention of the HC1 gas. Adherence to the procedures shown in the paper above provide the ability to easily use low level HC1 dry standards in the field. Therefore, it is not necessary, nor recommended, to run humidified HC1 calibration gas, at least up to the limits done for this testing. Humidification can add significant complexity to a calibration system and adds the possibility of condensation of water vapor which will cause increased response time due to HC1 absorption into the liquid phase. For best HCL test results PTFE calibration and heated sample lines should be used along with a dedicated Dursan - coated SS regulator that is never removed from the gas cylinder. Proper installation and pressure purging after the regular is installed will greatly reduce the chance of HC1 retention.
机译:从提供的数据可以清楚地看出,在HC1响应测试领域中遇到的问题应首先查看所使用的材料,测试气体的流量和调节器,这些都是造成HC1气体滞留的罪魁祸首。遵循以上论文中显示的步骤,可以轻松在现场使用低浓度的HC1干燥标准液。因此,没有必要,也不建议运行加湿的HC1校准气体,至少要达到该测试的极限。加湿会大大增加校准系统的复杂性,并增加水蒸气冷凝的可能性,这将由于HCl吸收到液相中而导致响应时间增加。为了获得最佳的HCL测试结果,应将PTFE标定和加热的样品管线与专用的Dursan涂层SS调节器一起使用,该调节器切勿从气瓶中取出。常规安装后正确的安装和压力吹扫将大大减少HC1保留的机会。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号