A quantitative analysis on the intracardiac electrogram contact during ventricular tachycardia ablation

机译:心室心动过速消融时心内电图接触的定量分析

获取原文

摘要

Cardiac arrhythmias are currently treated with ablation procedures, which are guided by X-ray cameras and intracardiac electrograms (EGM) recordings. Recently, the use of catheters with force sensors have been proposed for improving the contact of the sensing catheter over the heart wall, aiming to guarantee the reliability of electrical measurements. Manufacturers of the equipment used for this type of studies have suggested that the catheter has to take a reading above or equal to 5g. We aimed to determine whether the waveform of the EGM can provide us with some information about the good contact conditions of the catheter on the endocardium, by using the recorded force as a gold standard. We first performed a correlation analysis of EGM morphology in terms of the force threshold, and then we made a multivariable analysis based on the Fisher discriminant. The database to be analyzed came from 11 patients (8 males) with mean age of 64.9 years, where 1161 EGM signals were obtained. The EGM set corresponded to the voltage maps of the left ventricle during sinus rhythm. The correlation coefficient between consecutive EGM beats larger and lower than 0.8 in EGM morphology showed similar ratios of averaged force in those beats with larger force (49% vs 46% for force>5g), 29%o vs 28% for force<5g. Fisher discriminant analysis yielded error probability larger than 0.3 for the best discrimination case. These previous results show that the EGM morphology seems to have limited information about the catheter contact when scrutinized with simple signal processing methods.
机译:目前,心律失常是通过消融手术治疗的,该手术由X射线摄影机和心内电描记图(EGM)记录指导。近来,已经提出使用带有力传感器的导管来改善感测导管在心壁上的接触,旨在保证电测量的可靠性。用于此类研究的设备的制造商建议导管的读数必须大于或等于5g。我们的目的是通过使用记录的力作为金标准,来确定EGM的波形是否可以为我们提供一些有关导管与心内膜良好接触状态的信息。我们首先根据力阈值对EGM形态进行了相关分析,然后基于Fisher判别式进行了多变量分析。要分析的数据库来自11例患者(8例男性),平均年龄64.9岁,共获得1161个EGM信号。 EGM设置对应于窦性心律期间左心室的电压图。在EGM形态上大于和小于0.8的连续EGM搏动之间的相关系数显示,在具有较大力的搏动中,平均力的比率相近(49%vs.> 5g的力为46%); 29%o> <5g的力为28%。对于最佳判别情况,Fisher判别分析得出的错误概率大于0.3。这些先前的结果表明,当用简单的信号处理方法仔细检查时,EGM形态似乎对导管接触的信息有限。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号