CHARACTERIZATION OF PEEK AND ITS COMPOSITE COATINGS ON SUBSTRATE OF TITANIUM ALLOY FOR BIOIMPLANT APPLICATIONS

机译:钛合金基体上PEEK及其复合涂层的表征

获取原文
获取外文期刊封面目录资料

摘要

The tribological performances of PEEK composite coatings were better than that of pure PEEK coating. This has been attributed presence of Alumina and Silica nanoparticles, which were evenly distributed in PEEK coating. These nanoparticles prevented the asperities of upper samples from penetrating the surfaces of PEEK, and relieved the deformation of the surfaces as well as the plowing effect. Under the conditions used in the present study, PEEK/Al_2O_3 composite coating demonstrated optimum tribological properties and hence, could be taken as a potential candidate to be applied as bearing material for artificial cervical disc.
机译:PEEK复合涂层的摩擦学性能优于纯PEEK涂层。这归因于氧化铝和二氧化硅纳米粒子的存在,它们均匀分布在PEEK涂层中。这些纳米颗粒可防止上层样品的粗糙物穿透PEEK的表面,并减轻表面的变形以及耕作效果。在本研究条件下,PEEK / Al_2O_3复合涂层表现出最佳的摩擦学性能,因此可以作为潜在的候选材料用作人造颈椎间盘的轴承材料。