首页> 外文会议>IEEE Sensors >Flexible, structured MWCNT/PDMS sensor for chronic vascular access monitoring
【24h】

Flexible, structured MWCNT/PDMS sensor for chronic vascular access monitoring

机译:灵活的结构化MWCNT / PDMS传感器,用于慢性血管通路监测

获取原文

摘要

Graft wall pulsation amplitude sensing can provide a measure of functional status, e.g. in hemodialysis access grafts. Current implantable graft monitoring sensors require graft modification and direct bloodstream contact. We propose a new class of piezoresistive flexible pulsation sensors which can be wrapped around the graft to measure wall movement. These sensors must be highly flexible to prevent graft constriction; typical strain sensors are too rigid and the strain sensing range is too limited for this application. We describe a novel additive manufacturing (AM) method for printing polydimethylsiloxane (PDMS) with an internal porous structure, such that material compliance may be tuned anisotropically for a given sensor geometry. Prototype flexible pulsation sensors (FPS) consisting of structured PDMS with an embedded conductive PDMS sensor layer were fabricated and tested. Initial tests demonstrated reliable sensor response to 1-Hz cyclic elongation of 20%, and a sensor gauge factor of 1.0.
机译:移植壁脉动幅度感测可提供功能状态的量度,例如在血液透析中获取移植物。当前的可植入的移植物监测传感器需要移植物修饰和直接的血流接触。我们提出了一种新型的压阻柔性脉动传感器,可以将其包裹在移植物周围以测量壁的运动。这些传感器必须高度灵活,以防止移植物收缩。对于这种应用,典型的应变传感器太僵硬,应变感应范围也太有限。我们描述了一种新型的增材制造(AM)方法,用于印刷具有内部多孔结构的聚二甲基硅氧烷(PDMS),以便可以针对给定的传感器几何形状各向异性地调整材料的柔韧性。制作并测试了由结构化PDMS和嵌入式导电PDMS传感器层组成的原型柔性脉动传感器(FPS)。最初的测试表明,传感器对20 Hz的1-Hz循环伸长率具有可靠的响应,并且传感器的规格系数为1.0。