首页> 外文会议>IEEE International Ultrasonics Symposium >Real-time pulse compression in multigate spectral Doppler imaging
【24h】

Real-time pulse compression in multigate spectral Doppler imaging

机译:多门频谱多普勒成像中的实时脉冲压缩

获取原文

摘要

The quality of Doppler ultrasound blood flow investigations is often affected by a low signal-to-noise ratio due to the weak energy backscattered by erythrocytes flowing in deep vessels. Transmission of coded pulses and matched receive filtering improve the penetration depth while preserving the axial resolution. However, the involved additional computational load has probably so far discouraged the massive implementation of pulse compression methods in real-time Doppler imaging systems. In this paper, we report on the performance obtained by the pulse compression technique applied to (multigate) spectral Doppler imaging. A real-time algorithm was implemented on the ULA-OP research scanner. The results of phantom experiments show that up to 13dB signal-to-noise ratio gain can be obtained when a 5 μs-long linear chirp is transmitted. Furthermore, in-vivo tests demonstrate that pulse compression is an effective means to increase the penetration depth, by up to 2 cm, in spectral Doppler applications.
机译:多普勒超声血流检查的质量通常受信噪比低的影响,这是由于在深血管中流动的红细胞向后散射的能量很弱。编码脉冲的传输和匹配的接收滤波改善了穿透深度,同时保留了轴向分辨率。但是,到目前为止,所涉及的额外计算量可能不鼓励在实时多普勒成像系统中大规模实施脉冲压缩方法。在本文中,我们报告了通过应用于(多门)频谱多普勒成像的脉冲压缩技术获得的性能。在ULA-OP研究扫描仪上实施了实时算法。幻象实验的结果表明,当传输5μs长的线性线性调频脉冲时,可以获得高达13dB的信噪比增益。此外,体内试验表明,脉冲压缩是在频谱多普勒应用中增加穿透深度达2 cm的有效手段。